Calibration Parameter: Dew point, Frost point, RH%
Calibration Range: Dew point(-20—+70℃) |
Frost point(-25—0℃) |
RH(10 to 98% @ 0-70℃) |
RH(10 to 98% @ 0-70℃) |
Calibration Parameter: Dew point, Frost point, RH%
Traceable to: NIST / SI
Calibration Range: Dew point(-20—+70℃)
Frost point(-25—0℃)
RH(10 to 98% @ 0-70℃)
373LHX Dew Point Hygrometer
Specifications
Measuring Ranges
Frost/Dew Point: -75...95 °C
Temperature: -50...100 °C
Sample pressure: 500...2’000 hPa abs.
Accuracy
Frost/Dew point (for calibrated range): ≤±0.1°C
Temperature: ≤±0.07°C
Reproducibility
Frost/Dew point: ≤±0.05°C
Temperature: ≤±0.04°C
Standard Features
Digital I/O: RS-232
Mirror cooling: 3-stage Peltier thermoelectric
Mirror temperature sensors: Dual Platinum Resistance Thermometer PRT (Pt-100)
Analog outputs: Two analog outputs, user programmable, -10...+10 V and 4...20 mA
Gas connections: 6mm or 1/4’’ Swagelok
Sample gas flow rate: 0...1 l/min
Sample gas circuit: Stainless steel, 316 / 316L, electro polished
Cooling: Refrigeration
Internal pump: Yes
2500 Generator
Specifications
Relative Humidity Range:10 to 95 %RH
Relative Humidity Uncertainty: @ Pc Tc 0.5 %RH
Chamber Temperature Range: 0 to 70 oC
Chamber Temperature Range: (Optional) –10 to 70 oC
Chamber Temperature Uncertainty: 0.06 oC
Chamber Pressure Uncertainty: 0.15% FS
Chamber Pressure Range: Ambient
Gas Flow Rate Range: 5 to 20 L/min
Gas Flow Rate Resolution: 0.02 L/min
Gas Flow Rate Uncertainty: 2% FS
Gas Type: Air or Nitrogen
Gas Pressure Rating (MAWP): 175 psiG
Heating/Cooling Rate: 2.5 Minutes Per oC Average
Chamber Window: 6" x 6" (152 mm x 152 mm)
Air Supply (External): Clean Oil Free Instrument Air @ 175 psiG & 20 L/min
Applications
Used by NIMI, EPRI and academic institute for calibrating hygrometer
Used by dew point instrument manufacturers, humidity sensor and vapor content meter manufacturers for calibrating